BX8800 Full auto hematology analyzer
Specification sheet
Features:
Main Technical Specifications
Methodology: Electrical resistance for counting and SFT method of hemoglobin
Parameter: 3-part differentiation of WBC; 24 parameters include 3 histograms (WBC, RBC, PLT)
WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB,HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR,P-LCC;
Sample volume: venous 9.6uL, pre-diluted 20uL, capillary 9.6uL
Operation mode: Touch screen operation, also can be connected to mouse and keyboard
Throughput: More than 40 samples per hour, 24 hours work with auto standby
Storage: up to 100,000 sample results+3 histograms
Operation Language: English, French, Spanish, Portuguese, Russian, etc.
QC control: X-B, L-J, X, SD, CV %
Calibration mode: Manual calibration and automatic calibration
Display: 8.4 inches color touch screen
Reference: Male, Female, Children, Neonate
Printer: Thermal recorder, 57 mm width paper, various printout formats. Graphic thermal printer with various printing format, optional external printer
Input / output: 4×USB, 1×RS232,1×ethernet,1×VGA(for external monitor)
Temperature: 15℃- 35℃, humidity ≤10- 90%
Air Pressure: 70KPa ~ 106KPa
Power supply: AC100~240V 50/60Hz
Dimension: 33 CM (L) * 38 CM (W) * 43 CM (H)
Weight: 20 KG
Precision
Parameters linear range CV %
WBC( 10 9 / L ) 0.0 - 99.9 ≤2%
RBC ( 1012 / L ) 0.0 - 9.99 ≤1.5%
MCV ( fL ) 40-150 ≤0.5%
PLT ( 10 9 / L ) 0-999 ≤4.0%
HGB( g/L ) 0.0-300.0 ≤1.5%